Electrodeposited Anisotropic Pt Nanoarrays for Broad-Range Electrochemical Biosensing
Dong-Uk Kim1, Seung-Ki Baek1, Yong-Cheol Jeong2
1Strategic Planning Division, QuadMedicine, Inc. 45, Sagimakgolro, Hungwon-gu Seongnam, 13209, Republic of Korea
2Research Institute of Convergence Technology, KITECH 143, Hanggaulro, Sangnok-gu, Ansan, 15588, Republic of Korea
Correspondence:  Yong-Cheol Jeong, Tel: +82-31-8040-6197, Fax: +82-31-8040-6430, 
Email: ycjeong@kitech.re.kr
Received: 30 July 2026   • Accepted: 1 October 2026
Abstract
Hydrogen peroxide-mediated electrochemical biosensors often suffer from signal saturation at elevated analyte concentrations, which limits their usable linear range. Here, we report morphology-engineered platinum nanoarray (PtNA) biointerfaces formed on a commercial carbon screen-printed electrode (SPE) by controlled electrodeposition. The resulting electrode is hereafter denoted as PtNA/Carbon SPE. By tuning deposition potential and time, anisotropic Pt nanoarray-like architectures with increased electrochemically accessible surface area were obtained. Electrochemical impedance spectroscopy under the ferri/ferrocyanide probe condition indicated very low apparent charge-transfer resistance, while the amperometric results were consistent with a contribution from concentration-dependent utilization of accessible Pt regions within the PtNA architecture. The PtNA/Carbon SPE prepared under the selected fabrication condition showed a linear H₂O₂ response up to 2 mM and retained a larger fraction of its low-concentration sensitivity at elevated H₂O₂ concentrations than the commercial Pt SPE. After glucose oxidase immobilization, the PtNA-based biointerface maintained linear glucose detection from 0 to 7.5 mM without an additional diffusion-limiting layer, whereas the commercial Pt SPE saturated much earlier. These results show that deposition-controlled Pt nanoarray morphology can be a useful design parameter for enzymatic electrochemical sensing when broader linear operation is prioritized over ultralow detection limit.
Keywords: platinum nanoarrays, electrodeposition, enzymatic glucose sensor, hydrogen peroxide detection, screen-printed electrodes, electrochemical biosensing
TOOLS
Share :
Facebook Twitter Linked In Google+ Line it
METRICS Graph View
  • 0 Crossref
  •   Scopus
  • 337 View
  • 6 Download
Related articles in J. Electrochem. Sci. Technol

Effects of the Polarization Resistance on Cyclic Voltammograms for an Electrochemical-Chemical Reaction2015 December;6(4)



ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
E-mail: journal@kecs.or.kr    Tel: +82-2-568-9392               

Copyright © 2026 by The Korean Electrochemical Society.

Developed in M2PI

Close layer
prev next